Scan Acquisition |
Native |
Arterial |
Venous |
Range |
Above diaphragm to iliac crests |
Only liver lesion as identified by radiologist |
Repeat series every 15 seconds through lesion as directed by radiologist |
Care Dose 4D |
on |
on |
on |
Quality Reference |
200 mAs |
200 mAs |
200 mAs |
KV / effective mAs / rotation time |
120 / 200 / 0.5 |
120 / 200 / 0.5 |
120 / 200 / 0.5 |
Detector collimation |
1.5 mm |
1.5 mm |
1.5 mm |
Slice Thickness |
3.0 mm |
3.0 mm |
2.0 mm |
Pitch |
0.75 |
0.75 |
0.75 |
Kernel |
B30f |
B30f |
B30f |
Increment |
3.0 mm |
3.0 mm |
3.0 mm |
Image order |
Cranial to caudal |
Cranial to caudal |
Cranial to caudal |
Oral Contrast |
2 hour oral |
2 hour oral |
2 hour oral |
Intravenous Contrast |
None |
100 mL Isovue – 300 |
|
Injection Rate |
N/A |
3.0 cc/sec |
|
Scan Delay |
N/A |
45 seconds post injection. Call the Radiologist to locate liver lesion. Radiologist will then direct which area to scan in the intervals listed in the image order. DO NOT scan entire liver OR abdomen in serial interval scanning. |
Repeat serial scanning through liver lesion every 15 seconds for 3:00 minutes total. Lesion area selected by Radiologist. May perform 10 minute delay if requested. |
Reconstruction |
1. Recon 3x3 B30f mediastinum medium smooth 3. Recon 3x3 B80s 4. Recon 1x1 B30f 3D Card for coronal reconstruction |
6. Recon 3 x 3 B30f |
1. Recon 2 x 1 B30f |