| KV / effective mAs / rotation time |
120 / 200 / 0.5 seconds |
| Detector collimation |
0.75 mm |
| Slice Thickness |
1.0 mm |
| Pitch |
0.75 |
| Kernel (for initial reconstruction) |
B30f |
| Increment |
1.0 mm |
| Image order |
Cranial to caudal |
| Oral Contrast |
Water 250 cc every 30 min for total of 750 cc + 250 on table. (If emergent can give 750 cc H20 30 minutes prior + 250 on table) |
| Intravenous Contrast |
Isovue 300 - 120 cc. 18 - 22 ga IV needed preferably in the AC. |
| Injection Rate |
3.5 - 4 cc second |
| Scan Delay |
FIRST acquisition 30 - 35 seconds approximately. MUST USE CARE BOLUS. Place ROI on descending Aorta at level of the diaphragm. SECOND acquisition 70 second delay. |
| Reconstruction |
First acquisition reconstruction 1 x 1. First acquisition 2nd Recon and second acquisition B30f 3 x 3. |
| Additional Instructions |
Scan abdomen and pelvis. Perform thin MIP reformatting using 3D card: Image thickness 3, Distance between images 1, Number of slices 80 or whatever is needed to cover from anterior vertebral body to abdominal wall at L-2 level. |