Indications |
Evaluate Tracheomalacia or Tracheal stenosis |
Number of scan Acquisition/series |
3 |
Scan Acquisition |
1. Supine Inspiratory |
2. Dynamic Expiratory |
3. Repeat Dynamic Expiratory ( if necessary) |
Range |
2 cm above clavicular heads through both costophrenic angles |
2 cm above clavicular heads through main carina |
2 cm above clavicular heads through main carina |
Care Dose 4D |
On |
On |
On |
Quality Reference |
110 mAs |
40 mAs |
40 mAs |
KV / Effective mAs / Rotation time |
120 / 110 / 0.5 seconds |
120 / 40 / .42 seconds |
120 / 40 / .42 seconds |
Detector Collimation |
0.75 mm |
1.5 mm |
1.5 mm |
Slice Thickness |
1 mm |
2 mm |
2 mm |
Pitch factor |
1.3 |
1.5 |
1.5 |
Kernel (for initial reconstruction) |
B30f |
B80s |
B80s |
Increment |
1 mm |
1 mm |
1 mm |
Image order |
Cranial - Caudal |
Cranial - Caudal |
Cranial - Caudal |
Oral Contrast |
None |
None |
None |
IV contrast |
None |
None |
None |
Injection Rate |
N/A |
N/A |
N/A |
Scan Delay |
N/A |
Have patient forcibly exhale 1 second after technologist pushes button to initiate scan |
Have patient forcibly exhale 1 second after technologist pushes button to initiate scan |
Reconstruction (Series to send to PACS) |
1. Recon 3x3 B30f mediastinum medium smooth 2. Recon 2x2 B80f lung ultrasharp 3. Recon 1x1 B80s 4. Recon 1x1 B80s 3D Card for coronal reconstruction 5. MIP 20 mm slab 4mm increments 6. Recon 2 x1 B80s 7. Recon 2 x1 B80s |