KV / effective mAs / rotation time |
120 / 200 / 0.5 seconds |
Detector collimation |
1.5 mm |
Slice Thickness |
3.0 mm |
Pitch |
0.75 mm |
Kernel (for initial reconstruction) |
Abdomen B30f / Lung 80 / Bone B31f |
Increment |
3.0 mm |
Image order |
Cranial to caudad. Above diaphragm to the greater trochanters. |
Oral Contrast (Trauma Prep) |
None |
Intravenous Contrast |
100 cc Isovue 300 at 3cc/sec. 18-22ga Angiocath needed. |
Scan Delay |
75 sec - Abd |
Reconstruction |
First acquisition reconstruction 1 x 1. First acquisition 2nd Recon and second acquisition B30f 3 x 3. |
Additional Instructions |
Use thin 0.75 slice in 3D window to reconstruct Coronal and Sagittal reconstructions using 3 mm image thickness. |